200-240 V, 3+PE 50/60 Hz
-X01
Supply cable
380-415V, 3N+PE 50/60 Hz
-X01
Supply cable
NL3L2
L1
L3L2L1
-X01 terminal strip for incoming supply cable
Electrically-heated chamber and electrically-heated dryer
Voltage
[V]
Connection
Frequency
[Hz]
Max current
[A]
Fuse
[A]
380 3+PE 50/60 34 C40
400 3N+PE 50/60 35 C40
415 3N+PE 50/60 36 C40
Steam-heated chamber and electrically-heated dryer
Voltage
[V]
Connection
Frequency
[Hz]
Max current
[A]
Fuse
[A]
200 3+PE 50/60 25 C32
208 3+PE 50/60 24 C30
230 3+PE 50/60 22 C25
240 3+PE 50/60 21 C25
380 3+PE 50/60 14 C20
400 3N+PE 50/60 14 C20
415 3N+PE 50/60 14 C20