4. Screw the chamber extension module to the main module. Check that the seal is sitting
correctly before screwing on the chamber extension modules. Fill the gap between the modules
on the inner side of the chamber with silicone, the entire lower part and about 20 cm up the
walls.
• Fit the limit switch for closed door to each chamber extension module.
000525
• Adjust the chamber floor in relation to the floor level (tolerance 0 to +2 mm). The chamber floor
must be level with or slightly above the building's floor level. Use a straight-edge ruler and a
level. Check and adjust using the four machine feet and a crowbar.
000521