GETINGE-LA CALHENE ISOTEST | User manual
NTA
3015/78 C
Approuvé informatiquement / Electronic signature
DEX | Page 12/
74
WARNING
All maintenance operations must be performed after aerating
the circuits that contain hydrogen peroxide in gaseous form
and purging the circuits in which the sterilant is in liquid form.
It is compulsory to rinse with water when removing a
component on the injection circuit.
CAUTION
It is advisable to install hydrogen peroxide detectors in the
room where the isolator is installed (acceptable exposure
threshold: 1 ppm – 1.4 mg/m³).
It is advisable to install a luminous signal to show when a
biodecontamination phase is in progress to warn operators
who want to access the room (contact available on the
terminal block).
CAUTION
The outer door of the biodecontamination airlock SHOULD
NEVER be used as a support for setting any type of object,
even on a temporary basis. This may affect the adjustment of
the door, compromising its seal.
DANGER
Before opening the inner door, it is advisable to make sure
that no objects on the work top will interfere with its opening.
It is also imperative for the operator handling the door to
make sure that the second operator (using the other station)
has been warned and has not placed his hands under the
door. When opening, use one hand to hold the handle,
making sure to operate the door under optimum conditions.
Failure to observe this instruction may result in injuries to the
operator due to impact with the door or his hands getting
caught between the door and an object placed on the work
top.
CAUTION
When closing one of the two doors (inner door and outer door
of the biodecontamination airlock), it is advisable to check that
no objects will get caught between the door and the seal. This
may cause a leak-tightness fault. The leak test before a
biodecontamination cycle may fail in this case, making it
impossible to use the device.