8.7. RECIPE
The parameters are defined and validated by Getinge-La Calhène. Any modification may
cause a change in the cycle.
The "recipe menu" can be accessed from the main view with at least "validation" access
level.
WARNING
Do not modify the parameters of the "Isolator
biodecontamination", "Airlock biodecontamination", "Isolator
emergency" and "Emergency" recipes without first checking the
effectiveness of the aeration phase – See technical manual
(calibration and verification chapter).
Procedure for modifying recipe parameters
− In the user view, press the "recipe" button and the following view appears:
− Select the recipe to be modified.