7.6. HANDLING
Before each sterilization cycle, Getinge-La Calhène recommends checking the integrity of
the isolator gloves. The method consists of:
− conducting a test by raising the oxygen level to assess glove performance (GLT
testing device – operator manual NT 3020/1)
If the result of the test is "not correct", the glove must be replaced.
The procedure for replacing a glove without breach of containment is as follows:
Installation of gloves for an isolator under positive pressure
− Pull the glove
(2) to be replaced inside out into the sleeve (1).
− Remove the tightening ring (3).
− Position the new glove (5) parallel to the cuff ring (4) aligning the axis of the thumb
with the seal of the sleeve.
− Install the new glove (5) on the cuff ring (4) up from the bead (6) of the glove to be
replaced.
− Place the tightening ring (3) around the glove behind the bead (6).
− Position the bead (6) of the glove and the tightening ring (3) on the end of the cuff
ring groove
(4).
− Remove the glove (2) from the sleeve (1).
− Extract the glove (2).
1
2
3
4
1
2
3
4
6