DANGER
Premature halting of the aeration process will allow the
outer door to be opened with a residual level of H
2
O
2
in the
machine which is higher than the minimum level defined for
normal operation. This therefore represents a risk of
inhalation for the operator when opening the outer door of
the biodecontamination airlock.
If the "NO" button is pressed, the cycle continues.
The following view appears during production:
If the "YES" button is pressed, the cycle is halted and the isolator is no longer sterile.
If the "NO" button is pressed, the cycle continues.
8.4. EMERGENCY AERATION
In the event of a problem during the biodecontamination phase, the operator can launch
emergency aeration by pressing the red button next to the "EMERGENCY AERATION"
touch screen.