GETINGE-LA CALHENE ISOTEST | User manual
NTA
3015/78 C
Approuvé informatiquement / Electronic signature
DEX | Page 34/
74
Step 5: If the biodecontamination phase is completed with no errors, the isolator
automatically enters the waiting production phase and the airlock enters the
unloading phase.
Step 6: To start a production phase, press button the following
windows appears:
Step 7: Enter a batch number and press “Start new batch” button the production
phase starts automatically.
Perform the following operations
:
Close all the doors of the DPTE®.
Disconnect the multipurpose containers (or dummy containers).
Connect instead the necessary accessories (according to the chosen
option): DPTE-Tubing
TM
or DPTE-DispoBag
TM
.
Open the door of the airlock. Note that it will remain open for all the
remaining steps of the process.