Field Safety Notice' 2019-04-02 I 2018-025
Figure 1: Position mains connection on OR-Table / pulled out mains connection with exposed electronic
components/position of product label on mobile OR-Table MEERA including SN
Precautions
Until all actions have been accomplished, a visual and functional check is required to be performed by a trained
person before operation. Furthermore the information in the instructions for use (IFU) regarding connection to and
disconnecting from the mains socket (power source) should be observed as a matter of urgency. See figure 2
below.
Connecting to mains supply
GI Connect the supplied mains cable into the
mobile operating table (1).
E Then plug into the supply voltage socket.
.
7
The status display for mains operation
on the override control panel is lit up in
green
The rechargeable batteries of the mobile
operating table are charged.
.
7
The plug must always be accessible to
ensure that the OR table can be discon-
nected from the power supply at any
time.
Detach the mains cable
[RI Detach the plug at the main socket.
Cl
Detach the mains cable from the mobile op-
erating table.
Figure 2: Abstract form IFU (establishing / separating the mains connection)
If the plastic socket of the cold-device plug is pulled out do not try to put it back into the housing and if there are
any exposed electronic components: do not touch them. Call your in-house technical service and arrange the
repair of the table.