3031 Schüttelinkubator
Shaking Incubator
Seite / Page 23 Rev.02 vom / dated 31.03.2014
7. Technical data
Dimensions
Exterior dimensions (W x D x H)
525 x 665 x 570 mm
Cabinet dimensions (W x D x H)
450 x 450 x 280 mm
Volume / Capacity
approx. 46 litres, supports to accept one
shaking tray 3966
Net weight
38.5 kg
Temperature
Temperature range
+20°C (from approx. 8°C above ambient)
to +70°C,
cooling coil built in as standard
Temperature regulation
electronic, microprocessor-controlled,
PID type
Temperature setting and display
digital LED (0.1°C increments)
Temperature constancy (temporal)
+/- 0.2°C
Over-temperature cut-out
electronically settable up to 9.9°C above the
set value, to protect the test substances, and
electromechanical (approx 120°C) to protect
the unit
Under-temperature cut-out
electronically settable up to 9.9°C below the
set value, to protect the test substances
Shaking motion
Shaking motion type
orbital, can be switched on or off
Shaking frequency regulation
electronic, microprocessor-controlled, PID
type
Shaking frequency
10 - 250 min
-1
Shaking frequency setting and display
digital LED (1 min
-1
increments)
Shaking amplitude
30 mm
Load
Maximum load
12 kg
Timer
Incubation time
1 minute to 999 hours,
microprocessor-controlled
Time presetting and display of remaining
running time
digital LED (hours or minutes)
Interface
RS232
Electrical connection
Mains voltage and mains frequency
230V or 115V, +/-10%, 50....60Hz
Power
800 W
Mains fuse, user's place
10 A
Protection type / protection class
I / IP20
Environmental conditions
only for indoors use
(not in explosion endangered surroundings!)
Temperature
+10°C to +30°C
Humidity
maximum 70% relative humidity,
non-condensing