LPS10
- 14 -
V1.0
Units of measurement:
Changing the temperature units of measurement affects the Modbus protocol only. The
temperature measurement provided by the ASCII proprietary protocol (commands
S0/S1) is always in °C.
Command Reply Description
TTn WUT;&| Sets the temperature unit of measurement:
°C if n=0 (default)
°F if n=1
K if n=2
RUT RUT;n| Reads the temperature unit of measurement.
Measurements reading:
Command Reply Description
S0 S0;meas| Prints the measurements in the following sequence:
Temperature compensated irradiance in W/m
2
(it takes into account the change in sensor sen-
sitivity as temperature changes)
Nominal irradiance in W/m
2
(calculated considering the sensor nominal sen-
sitivity at the calibration temperature:
23 °C)
Signal in mV generated by the thermopile
Internal temperature in °C
Internal relative humidity in %
Internal pressure in hPa
Tilt angle (only LPS10MxT)
Power supply voltage
If the automatic sending of the measurements is enabled
(command S1), sending S0 disables the automatic send-
ing.
S1 S1;meas| Enables the automatic sending of the measurements, once
per second.
Diagnostics alarm thresholds:
Command Reply Description
WAL1n…n WAL1;&| Sets to n…n the operating time alarm threshold fro
m first
power on (0…65535 days, default=730 days 2 years).
AL1 AL1;n…n| Reads the operating time alarm threshold from first
power on.
WAL2n…n WAL2;&| Sets to n…n the operating time alarm threshold from
last power on (0…65535 days, default=730 days 2
years).
AL2 AL2;n…n| Reads the operating time alarm threshold from last
power on.
WAL3n…n WAL3;&| Sets to n…n, in the set unit of measurement, the inter-
nal temperature alarm threshold (0…150 °C, default=80
°C).
AL3 AL3;n…n| Reads the internal temperature alarm threshold.
WAL4n…n WAL4;&| Sets to n…n% the internal relative humidity alarm
threshold (0…100%, default=25%).
AL4 AL4;n…n| Reads the internal relative humidity alarm threshold.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GHM Delta OHM PYRAsense LPS10 Series and is the answer not in the manual?

GHM Delta OHM PYRAsense LPS10 Series Specifications

General IconGeneral
BrandGHM
ModelDelta OHM PYRAsense LPS10 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals