35
3.2.2 Delete Data
On menu screen, select “Delete Data” and press “¢” key to enter into Deletion
Conrmation screen. Then, select “Yes” with “p” key (short time press), after
short pressing “¢” key, all the records will be deleted.
Figure 14 Delete Records
3.2.3 Over-limit Alert Settings
On menu screen, select “Alert” and press “¢” key to enter Alert Limit Settings
screen, as shown in Figure 15.
Figure 15 Over-limit Alert Settings
• SpO2 Lo: set SpO2 low limit; Setting range:85%~100%, default setting:
85%;
• PR Lo: set pulse rate low limit; Setting range:25bpm~99bpm, default setting:
50bpm;
• PR Hi: set pulse rate high limit; Setting range: 100bpm~250bpm, default set-
ting: 120bpm;
• Alert: enable/disable the function of over-limit alert by audible beeping, visual
display and vibration of the oximeter.