by measurement and nd the hypoxemia patient earlier, thereby preventing or
reducing accidental death caused by hypoxia effectively.
3 Factors affecting SpO2 measuring accuracy (interference reason)
Intravascular dyes such as indocyanine green or methylene blue.
Exposure to excessive illumination, such as surgical lamps, bilirubin lamps,
uorescent lights, infrared heating lamps, or direct sunlight.
Vascular dyes or external used color-up product such as nail enamel or color
skin care.
Excessive patient movement.
Placement of a sensor on an extremity with a blood pressure cuff, arterial
catheter, or intravascular line.
Exposure to the chamber with High pressure oxygen.
There is an arterial occlusion proximal to the sensor.
Blood vessel contraction caused by peripheral vessel hyperkinesias or body
temperature decreasing.
4 Factors causing low SpO2 Measuring value (pathology reason)
Hypoxemia disease, functional lack of HbO2.
Pigmentation or abnormal oxyhemoglobin level.
Abnormal oxyhemoglobin variation.
Methemoglobin disease.
Sulfhemoglobinemia or arterial occlusion exists near sensor.
Obvious venous pulsations.
Peripheral arterial pulsation becomes weak.
Peripheral blood supply is not enough.