67
Guidance and manufactures declaration-electromagnetic immunity
The device is intended for use in the environment specified below. Buyer or operator should
assure that it is used in such environment.
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic environment -guidance
Conducted
RF
IEC61000-4
-3
3 Vrms
150kHz
80MHz
3 Vrms
Portable and mobile RF communication
s
equipment should be used no closer to any
part of the device, inc
luding cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d=1.2√P
d=1.2√P 80MHz-800MHz
d=2.3√P 800MHz-2.5GHz
Where P is the maximum output power ra
ting
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
det
ermined by an electromagnetic site survey,
a should be less than th
e compliance level in
each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF
IEC61000-4
-3
3 V/m
80 MHz
2.5 GHz
3 V/m
NOTE 1: At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the device or system is
used exceeds the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as adjusting the direction or location of the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.