If the current wave is good, please click the "Accept" button, and then the system enters the
arrhythmia analysis (as Fig.6-6) automatically. If the user wants to exit this program, please
close direct. If the current wave is not good, please click the "Next Wave" or "Next Segment"
button, then the system will show the waves constantly until you click "Accept" to enter the
arrhythmia analysis.
Click the "▼" button at the right of "Show Lead" to choose other leads as main analytical lead.
RR refractory period: This parameter is general 300ms, it means the shortest time between the
two heart beats, the default value is 300ms, the user can adjust it according to the specific
circumstance, if the patient's heart rate is too quick, it should be set as low as possible, in order to
prevent losing the analysis for some heart beats.
Click the options under the "Leads Analyzed" could decide which leads to be analyzed, the
default is 8 collected leads.
When the amplitude of the case's wave is too low, please choose "H" in the resolution option
("Height").
When the case meets much disturbance, please choose "H" in the "distinguish O" Option. The
options "Height" and "distinguish O" don't need to adjust generally. The user can choose
according to the actual circumstances.