Bravo pH Monitoring System 12 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
c. Press
Enter
. You move on to the first screen of the next setting:
Show pH
value
of
Preferences
.
4. The Preferences screen allows you to define setting that affect all studies.
Show pH Value: If
Yes
, the current pH value appears on the recorder screen
during studies. If
No
, pH values are only displayed for the first 30 minutes of a
study. The factory default is
No
.
Button Beep: If
Yes
, the recorder beeps when the patient presses a symptom
button. The factory default is
Yes
.
Capsule LED Blink: If
Yes
, the recorder capsule LED blinks when the capsule
signals are received. The factory default is
Yes
.
Set Date/Time: Once set, the recorder maintains the correct date and time. You
should only have to change this again if there is a time change (for example,
going on or off of Daylight Savings), or if the recorder battery is allowed to fully
discharge.
Date Format: You can set the date format to
MM-DD-YY
or
DD-MM-YY
. The
factory default is
MM-DD-YY
.
Time Format: You can set the time format to
12-hour (AM/PM)
or
24-hour
(military)
. The factory default is
12-hour
.
PC Software: Select the software used to record, view, and analyze the data
from studies. The choices are
AccuView
(factory default applies to Reflux
software, as well) and
RAPID pH
.
Language: Select the language for the recorder interface. The choices are
English, Danish, Dutch, Finnish, French, German, Italian, Norwegian,
Portuguese, Spanish, and Swedish. The factory default is
English
.
Backlight Duration: Set the time that the recorder screen backlight stays on
after a button is pressed. The choices are 15, 30 (factory default), 45, or 60
seconds.
5. To return one level up in the menu tree, press
ESC
.
֠
Note
At any time you may select Settings from the Main screen and
access Preferences to review and set the parameters.
֠
Note
Once you have set preferences, you will not need to reset them
unless:
you want to change something, or
the recorder battery is fully discharged.