Bravo pH Monitoring System 25 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
11. At the clinician’s discretion, endoscopically confirm the capsule’s attachment.
12. Confirm that the recorder is recording pH values and that the capsule status LED is
blinking in blue.
13. Proceed to Reviewing Instructions with Patients on page 26.
Stopping a pH Study
The study completes automatically when the recorder no longer collects pH data and the
recorder screen turns off. The study data is stored in the recorder until it is cleared.
To stop a study manually (before its intended completion):
1. Press the on/off button for several seconds until this message is displayed:
Are you
sure you want to stop the pH study?
2. Select your action:
• If you select
No
, the study continues uninterrupted.
• If you select
Yes
, this message is displayed:
Last study data not uploaded! To
upload, connect recorder to PC.
Remember: if you stop the study, you must upload the data to unlock the recorder
before you can continue.
Caution
Avoid contacting the capsule with the endoscope. Contact between the
endoscope and capsule may result in the dislodgement of the capsule.
Note
The recorder's display turns off after 60 seconds to conserve battery
life. Press any key to activate the recorder's screen during a study.