Bravo pH Monitoring System 24 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
If the plunger does not automatically spring up, push it up gently with your thumb.
These steps release the capsule from the delivery device.
Figure 12. Release the capsule from the delivery device, cont.
.
.
10. Turn off the vacuum source. Remove the delivery device and discard it according to
local waste management regulations.
!
Warning
Do not rotate or otherwise force the plunger beyond the white line on
the barrel. Rotating or forcing the plunger beyond this may result in
possible damage to the delivery device. This may also interfere with
the detachment of the capsule from the delivery device, and cause
possible injury to the patient.
֠
Note
Following correct rotation of the plunger, the white marking should now
be visible on the sixth rib of the plunger. If not, use your thumb to raise
the plunger until the white marking is visible.
!
Warning
If problems occur with capsule detachment, see Delivery Device
Disassembly Procedure on page 30.
!
Warning
Do not remove the delivery device from the patient with the vacuum
source on. Removing the device with the vacuum source on can result
in dislodgement of the capsule or possible injury to the patient’s
esophagus.
Do not cover the hole
at the bottom of the
handle with your
hand.
The plunger should spring back up
automatically. Verify that the
plunger springs back to the white
line marking the plunger’s sixth rib.
If the plunger does not automatically
spring up, use your thumb to lightly
push it up.
a
b