Bravo pH Monitoring System 20 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
5. Turn off the vacuum and detach the tubing from the delivery device.
6. Proceed to Starting Recording.
Starting Recording
1. If the recorder is turned off or in sleep mode, turn it on.
2. From the main menu, select
Start Study
and press
Enter
.
If the recorder detects a conflict with the capsule ID, this message appears:
Capsule
ID mismatch. Use pH capsule ID: xxxxx.
See Recorder Troubleshooting on
page 32.
3. Verify that the recorder is recording pH values.
4. Proceed to Placing the Capsule on page 21.
Caution
If the vacuum readings fail these required
minimum levels, remove the delivery device
and perform the same action on the vacuum
tube alone:
• Cover the tube with your finger. The
vacuum gauge should read at least
575 mmHg. Remove your finger. The
vacuum gauge should drop drastically (to
zero or close to zero). If so, the vacuum
unit is functioning properly. Perform the
procedure with a different delivery device
(after performing calibration).
• If the vacuum fails to reach 575 mmHg or
fails to drop to close to zero, the vacuum
unit may require service. Contact customer support.
Note
The recorder screen shuts off after 60 seconds to conserve battery
power. To turn on the recorder screen during a study, press any button.