Bravo pH Monitoring System 23 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
8. Swiftly press the plunger on top of the handle all the way down until it stops at its
locking position. This advances the trocar needle into the suction chamber
(Figure 10).
Figure 10. Attach the capsule to the patient’s esophagus.
9. Using your thumb, rotate the plunger from the side one-eighth (1/8) of a turn
clockwise to release the capsule from the delivery device (Figure 11). The plunger
springs up so that a white line is visible on the sixth rib of the plunger (Figure 12).
Figure 11. Release the capsule from the delivery device.
!
Warning
Press down on the plunger with a swift and smooth motion to actuate
the delivery device mechanism. Pressing down on the plunger too
slowly may result in the capsule not properly attaching to the patient’s
esophagus or not detaching from the delivery device.
Do not rotate the plunger while depressing it!
֠
Note
Your hand may feel the delivery mechanism actuate, and you may
hear a click when this occurs.
Remove your thumb after the plunger locks and before you begin
rotating it 1/8 of a turn.
Press the plunger down to the locking position and release it.
Rotate the plunger clockwise until the white line on the plunger lines up with the white upward
arrow.