At the end of this calibration period, the recorder beeps and this message appears:
pH capsule ID xxxx calibrated.
(where
xxxx
is the capsule ID).
18. Press
OK
. The calibration data is saved and the recorder main menu appears with the
cursor ready at
Start study
.
19. Proceed to Performing a Bravo pH Study on page 19.
Caution
If either part of the calibration process (pH 7.01 or 1.07) fails, see
Recorder Troubleshooting on page 32.
Note
If you are performing a two-capsule procedure, the recorder guides you
through the same calibration process for the second capsule.
After the first capsule is calibrated, it can remain outside the buffer
solutions for up to 60 minutes before use.
Note
If you are not going to place the capsule in the patient immediately,
return the capsule to the pH 7.01 buffer solution. It may remain there
for up to 8 hours. Rinse the capsule in sterile water before use.