Bravo pH Monitoring System 33 User Guide
chapter.fm 3/21/16 4:05 pm
5.5 x 7.75 inches (140 mm x 197 mm)
Recorder displays message:
Listening for caps... and does
not progress further.
Capsule out-of-range. Try each of these:
1. Move recorder closer to
capsule.
2. Replace buffer solution.
3. Try calibrating a new pH
capsule.
If error persists, contact
customer support.
Start Study
Recorder displays message:
Charge battery.
Recorder battery is below 90%
capacity.
Connect recorder to charger.
Allow it to fully recharge (may
take up to 7 hours).
Recorder displays message:
Settings for two capsules but
only capsule xxxx found. Start
recording?
Recorder was configured for
two-capsule study, but is now
only detecting signal of one
capsule.
If you continue recording, you
will only collect data for the
capsule identified in this error
message.
Otherwise, press
ESC
,
perform calibration with
another capsule, and continue.
Recorder displays message:
Capsule ID mismatch.
Use capsule ID: xxxx.
Capsule was not calibrated
using this recorder, or there
may be interference with
another capsule in the area.
Try each of these:
1. Select the recorder that
was calibrated to the
capsule being used
2. Move recorder closer to
capsule (place recorder on
patient’s chest).
3. Remove patient from the
area (100 meters) and try
again.
Recorder displays message:
Capsule ID mismatch.
Use capsule ID: 0000.
Calibration data is missing or
invalid.
Recalibrate the capsule. Make
sure that you get the message
pH capsule #1 calibrated.
No pH display on recorder screen Capsule out-of-range. Move recorder closer to
capsule.
Capsule not in contact with
fluid.
Place capsule in buffer
solution.
Data Transfer
Recorder displays the
do not
disconnect
icon
Recorder is connected to PC
software application and data
is being transferred.
This is normal. Do nothing (do
not disconnect recorder from
USB cable until this icon is no
longer displayed, indicating
that data transfer has finished).
Data upload fails. Communication error. Disconnect the recorder from
the USB cable and then
reconnect it.
Problem Cause Solution