EasyManua.ls Logo

Gmate Mini - Page 35

Gmate Mini
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
Performance Characteristics
Within run precision(100 venous blood tests per glucose level)
Mean glucose
Standard
deviation(mg/dL)
Coefficient of
variation(%)
41 mg/dL(2.3 mmol/L)
102 mg/dL(5.7 mmol/L)
147 mg/dL(8.2 mmol/L)
237 mg/dL(13.2 mmol/L)
394 mg/dL(21.9 mmol/L)
1.2
2.3
3.7
4.7
8.6
2.8
2.3
2.5
2.0
2.2
Gmate
TM
mini System
Total precision(100 control solution tests per glucose level)
Mean glucose
Standard
deviation(mg/dL)
Coefficient of
variation(%)
40 mg/dL(2.2 mmol/L)
106 mg/dL(5.9 mmol/L)
322 mg/dL(17.9 mmol/L)
0.9
2.3
6.4
2.2
2.2
2.0
Gmate
TM
mini System

Related product manuals