GS700 USER GUIDE
TWA Alarms
A Time Weighted Average (TWA) gas level, is the
average gas level measured over a speci c period
of time.
» Short Term Exposure Limit (STEL) - 15 minutes
» Long Term Exposure Limit (LTEL) - 8 hours
Toxic gases have STEL / LTEL alarm set-points.
The examples below show:
» H
2
S STEL alarm, with a gas reading of 12PPM and a STEL alarm  ag.
» CO LTEL alarm, with a gas reading of 38PPM and a LTEL alarm  ag.
Alternates with
Alternates with
i
Live readings are shown (not calculated averages).
Over Range Alarm
In the event of the  ammable gas (LEL) sensor
being exposed to a high concentration of  ammable
gas, the instrument has an over- range alarm.
If the LEL sensor is exposed to a gas reading above
100% LEL, EEE will alternate with the HIHI alarm
ag.
i
The ppm range will display the over- range alarm,
EEEE, when the gas reading is above 10,000ppm.
Only when con gured not to auto-range to LEL.
Alternates with