42
GS700 USER GUIDE
Instrument Speci cation
Dimensions
190 x 98 x 107 mm
Weight
1.4 kg (Alkaline), 1.3 kg (Rechargeable)
Operating Temperature
-20
o
C to 50
o
C (-4
o
F to 122
o
F)
Humidity
0 - 90% RH non-condensing
Construction
Polycarbonate ABS with TPE over-moulding
IP Rating
IP55 (Protected from dust & water jets)
Display
Monochrome LCD (240 x 160 pixels) with automatic bi-colour
backlighting
Alarms
360
o
highly visible  ashing LED
Sounder ~ 85 db at 0.3 m
Sampling System
Integral pump with  ow fail sensor. Sample path is protected
by a hydrophobic and cotton  lter.
Flow rate: ~ 0.5 l/min with no restriction.
Response times increase ~ 2 second per metre of tubing
used. Maximum tubing length 30 metres.
Power Source
Alkaline Batteries: 3 x size ‘D’ (LR20), or
Rechargeable Battery Pack
Only use approved alkaline batteries.
Approved Batteries
Alkaline: DURACELL: PROCELL (MN1300) or
INDUSTRIAL (ID1300)
ANSMANN: INDUSTRIAL or X-POWER
PANASONIC: EVOLTA
Rechargeable: GMI rechargeable battery pack
Battery Life
Alkaline: 15 hours
Rechargeable: 20 hours
Warranty
2 years (excluding consumables)