Maintenance
Cleaning
Display Window
• An optical cloth MUST be used to clean the display window, as the window is susceptible
to scratches. In extreme cases a mild screen solution may be used with the optical cloth
to remove stubborn stains.
Outer Casing
The outer, impact resistant, casing of the GS700 may be cleaned using a non-abrasive
damp cloth. Rub the cloth over the outer casing to remove any dirt and grime. In extreme
cases, a mild soap solution may be used with a non-abrasive cloth to remove stubborn
marks.
!
Do not use polishes containing silicon or solvent to clean the instrument as these may
affect the gas sensors.
!
Do not use abrasive materials or strong volatile chemical solutions as these could damage
the impact resistant casing.
Probe Handle Filter Replacement
Hydrophobic and cotton particulate  lters in the probe handle minimise the danger of water and
dust ingress.
If a blockage occurs, the message ‘SAMPLE FAULT’ will be displayed. Check the sample line and
probe handle for a blockage. When cleared, press
B1 to continue.
Replace the  lter(s) if the message does not clear.
To replace the  lter(s):
1. Unscrew the probe handle assembly.
2. Remove the cotton particulate  lter and
discard.
3. Remove the hydrophobic  lter.
4. Clean the probe handle to make sure it’s
free from dirt and water.
5. Fit a new cotton particulate  lter.
6. Fit the new hydrophobic  lter. The yellow
label on the  lter ts against the yellow
label on the probe handle.
7. Reassemble the probe handle assembly.
Hydrophobic Filter
Cotton Particulate Filter
Probe Handle Assembly