37
MAINTENANCE
Replacing the Dust Filter
Cotton particulate (dust)  lters, minimise the danger of dust ingress.
Removing the inlet nozzle allows access to the dust  lter.
1. Using a small coin or appropriate sized flat bladed screwdriver, remove the inlet nozzle.
2. Remove the dust  lter from the inlet nozzle.
i
The dust  lter should be removed for inspection periodically.
3. Clean the inlet nozzle to make sure it’s free from dirt and water.
4. Fit a new dust  lter into the inlet nozzle.
5. Fit the inlet nozzle.
i
Ensure the inlet nozzle screw thread is correctly located, before  tting.
!
The GS700 must never be switched on without suitable  lters installed.
Inlet NozzleDust Filter