Fault Alarm Types
Sample Fault
A fault with the GS700  ow due to the sample path being
blocked, water ingress, a blocked  lter or pump failure.
Functionality is disabled and ‘Sample fault’  ashes
OFF and ON.
Audible alarm, visual red LED’s on top plate and red
display backlight activate.
The sample path and  lters should be checked for
a blockage or water ingress.
Once the blockage is cleared, press
B1 to restart
the pump.
i
In CSM mode, once the blockage is cleared, the
pump will automatically restart.
Zero Fault
Indication that the zero gas reading is outwith its calibration
limits.
Zero alarm  ag and %LEL alternate on the display.
Audible alarm , visual red 360
o
LED bar activate.
• Press
B3, to zero.
i
Always Zero in fresh air.
Alternates with
Alternates with