10 GOSSEN METRAWATT GMBH
7 Current Source and Current Sink
Connect the device under test with the measurement ca-
bles.
Select the mA current sink calibration function or
mA current source with the rotary switch.
Press Esc until you enter the range selection menu.
Select the desired range:,
0 20 mA, 4 20 mA or 0 24 mA,
with the #" keys.
Press the key to enter the output menu.
Setting the simulation value:
The decade (i.e. the position of the digit to be changed) is
selected with the $ ! keys, and the respective digit is
set with the #" keys.
ON indicates that the current source is active.
7.1 Current Sink – Simulation of a 2-Wire Transmitter
A current sink (0 24 mA) or current loop load can be simu-
lated with this function. The calibrator regulates the current,
which flows via the calibrator jacks from an external power
supply, independent of direct voltage applied to the jacks
(4 27 V). The caliibrator varies the internal resistance such
that the adjusted current value flows.
Note!
The latest adjusted simulator range is stored to mem-
ory. Voltage at the calibrator jacks may not exceed
27 V in the current sink operating mode, because
thermal overload would otherwise occur and the fuse
F3 would blow.
LoVolt appears at the display where the voltage is too
low.
Example of a 2-Wire Transmitter Measuring Circuit
7.2 Current Source
Internal supply power is used for the simulation of a current
source.
Note!
The internal control circuit of the current source is
monitored: if voltage drops by more than 20 V at the
external burden or if the current circuit is interrupted,
„Hi burd“ appears at the display.
Attention!
!
In this operating mode interference voltages must not
be applied to the calibrator jacks.
In the event of operating errors, such as the short-
term application of a high interference voltage, the in-
strument/calibrator is protected by a replaceable
fuse F3, see chapter 12.3.
Simulation in Percentages (current source only)
Select the simulation function and the output range
(Isink/Isource = 4 20 mA) with the rotary switch and
the # key.
Press the ESC and # keys simultaneously.
13.452
NUM ON
SINK
mA
4-20
mA
4-20
NUM
sele t range
mA
#"
Current
#"
Current Value
Range
$ !
ON
SINK
SINK
Range Selection Menu
ESC
(... 2x)
Peripheral Device
24 V
Power
4 ... 20 mA = 0 ... 100 °C
+
+
Pack
CALIBRATOR
+
Output
13.452
NUM ON
SOURCE
mA
0-20
mA
0-20
NUM
sele t range
mA
#"
Current
#"
Current Value
Range
$ !
ON
SOURCE
Range Selection Menu
ESC
(... 2x)