12 GOSSEN METRAWATT GMBH
Automatic Interval Sequence
Automatic execution of a programmed sequence range is
above all advisable if feeding to a signal circuit, and scanning
of the peripheral device under test are physically separated.
After configuring all parameters in accordance with the menu
flowchart on page 13 for the “automatic interval sequence”
(Int, Auto = yes), the sequence can be started, and stopped
or continued at any desired time.
Example of an Automatic Interval Sequence
Interval parameters:
Output quantity: U (0 15 V range),
Start =2V, End = 10 V, number of interval steps = 3, t1 = 5 s,
auto = Yes (yes for automatic sequence)
Key
:
1 The sequence is started by pressing the key when Int
start is displayed (see Menu Flowchart on page 12).
2 The sequence is stopped by pressing the " or the #
key. Interval time elapsed thus far is saved as t
x
.
3 The sequence is resumed by pressing the " key, and
remaining sequence duration t
y
is equal to t1 t
x
.
4 Stop the interval sequence by simultaneously pressing
and holding the " and the # key (2 audible acoustic sig-
nals must be generated).
Menu Flowchart: Starting and Stopping the Interval Sequence
2
4.6
7,3
10
t[s]
U
[V]
run up stop
➀➃
510152025
stop
0
run dn hold run up run dn
continue
LCD:
+
"
#"
#
"
Int
set start
Int
ready
05.000
0-20 run dn
#
"
+
#"
#"
Start must
2x
mA
mA
Start
INT ON
SOURCE
#"
sequence.
blink.
#
"
+
2x
Stop the
interval sequence.
#
"
1x: Stop sequence: Display: hold (hold)
2x: Continue sequence: Display: run dn (run down)
1x: Stop sequence: Display: hold (hold)
2x: Continue sequence: Display: run
Up (run up)
ESC during the sequence (run or hold):
Return to calibration function.
ESC when stopped:
Return to sequence.
Int
stop
ESC
ESC