Cleaning and Disinfection of the Pump
WARNING: Ensure pump power is off and unplug the power cord from the wall outlet
before cleaning the pump.
s NOTICE:Donotoodanypartofthepumpwithcleaningsolution.Donotimmersethe
pump in liquid.
s NOTICE:Avoidspillinganyliquidonpump.Ifspillsdooccur,cleanuidfrompump
wearing rubber gloves or while unit is unplugged to avoid any possibility of shock. Once
liquid is removed, check operation of components in area of spill. Any liquid remaining
on pump can cause corrosion, which may cause components to fail or operate erratically,
possibly producing potential hazards for patient and staff.
Clean the pump weekly using a clean, damp soft cloth and mild detergent.
1. Wipe all controls, chassis and hose ttings with a damp cloth and a mild detergent.
2. Using a clean nylon brush, gently clean all crevices, as they can harbor micro organisms.
3. Air dry all treated surfaces.
The pump casing is manufactured from ABS plastic; if the case is soiled, the pump can be wiped
down with a sodium hypochlorite solution to dilution of 1000 ppm or any EPA-approved hospital
grade disinfectant (do not use phenol base cleaning solution).
INSPECTION / SYSTEM CHECK-OUT
Check each of the following before placing the Mattress System with a new patient:
1. Check mattress surface for tears or cracking; do not use if tears or cracks are present.
2. Ensure mattress is free of stains and is not overly faded.
3. Ensure air inlet hoses and connectors on mattress and pump are clean and undamaged.
4. Ensure pump and power cord are clean and undamaged.
5. Ensure pump hanger brackets are secure and operate correctly.
6. Ensure ON / OFF Power switch and comfort control knob both operate correctly.
7. Attach pump to Alternating Pressure hoses and power up to ensure there are no air leaks.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Graham Field LUMEX SELECT Series and is the answer not in the manual?

Graham Field LUMEX SELECT Series Specifications

General IconGeneral
BrandGraham Field
ModelLUMEX SELECT Series
CategoryIndoor Furnishing
LanguageEnglish

Related product manuals