LS300-INS-LAB-RevD19 16
8 TROUBLESHOOTING
The following list of problems, their causes and solutions will assist you in determining what may
be causing your Select Air Mattress not to function as designed. If a problem occurs which is not
listed below, contact GF Tech Support at 1.770.368.4700 for further information. Do not attempt to
repair components or parts, as this may invalidate your warranty or cause further problems that
may result in patient injury. Stop using your mattress immediately if it is not functioning correctly
or any warning beeps are heard.
If any of the following notications occurs, follow the steps below to troubleshoot:
Review all selections of this manual before troubleshooting any Select Mattress System.
Do not attempt any troubleshooting not shown in this manual or where the remedy recommends
contacting a GF authorized distributor. Any unauthorized service, modication, alteration, or
misuse may lead to serious injury and / or product damage and will void all applicable warranties.
SYMPTOM POSSIBLE CAUSE SOLUTION
Air is pumping out from the control unit
but mattress is not inating
Faulty power source — improper
voltage may cause the pump to
function abnormally and damage the
control unit
Use a power regulator
Kinks in the air tubes Adjust the air tubes to enable smooth
air ow
Leakage from the air cells Replace air cell if faulty
Leakage from air tube between
mattress and pump
Replace with new air tubes
Improper air tube connection Re-connect the air tubes
The pump is not functioning Power cord or power voltage Use a power regulator
Faulty fuse Replace the fuse
Some of the air cells are not properly
inated
Kinked connection between air cells
and manifold
Correct kinking between air cells and
manifold
Leakage from the air cells Replace air cell if faulty
WARNING:
If you experience a problem with your mattress system and are unable to service
it yourself, contact GF Tech Support at 1.770.368.4700 or your GF authorized distributor.