16
1. Press the key to access the setting up.
2. Use the key to select the desired mode.
3. Press the key to confirm.
4. To continue with other setups use the ► key.
5. To exit menu and return back to weighing mode, press the key.
UF-9 Standard Gravitational Pre-calibration (G Value)
Even the scale allows doing the Standard Calibration, Gravitational Calibration can
help to provide higher accuracy to the scale. This pre-calibration is allowed when the
switch CAL is at ADJ position. Gravitational value will be saved at -00- and will be
replaced each time a new value has been entered.
Sender G value: set it before Weight Calibration
Recipient G value: set if after Weight Calibration
The G value will be denied when the value is greater than 9.83217 (Polar G value) or
less than 9.78031 (Equator G value)
Factory Default: 9.79423
1. Press the key to setup.
2. Le display will show the last value for 1 second.
3. Use ◄, ► keys to see previous entered values. The display will stay at 00 if no
value has been entered.
4. The display will show the last value for 1 second.
5. Use ◄, ► keys to see previous entered values. The display will stay at 00 if no
value has been entered.
6. Press the key.
7. Use ◄, ►, ▲, 0 ~ 9 keys to enter the desired value.
8. Press the key.
9. Press the key to continue with other setups or the key to exit and return
back to normal weighing mode.