OCF 5.0 Open Channel Flow Monitor
SIMULATION
Exercises the 4-20mA (0-5V) output, digital display and control relays.
Test Select Maximum and press to simulate maximum Flow, Range or
Level and to output 20mA (5V) to the analog channel.
Select Minimum and press to simulate minimum Flow, Range or
Level and to output 4mA (0V) to the analog channel.
To simulate an intermediate Flow, Range or Level set Test to Actual
and then enter a value for the Flow, Range or Level. The analog
output and control relays will respond to the simulated value.