5
Connection terminals for inputs and outputs
Starting up the product
WARNING
Electric shock
Death or serious
personal injury
- A supply disconnecting
device made according
to EN 60204-1 5.3.2
must be provided with
means permitting it to
be locked in the OFF
(isolated) position. This
device must also be
installed in a position
according to EN
60204-1, 5.3.4.
WARNING
Electric shock
Death or serious
personal injury
- Power cables without a
plug must be
connected to a supply
disconnecting device
incorporated in the
fixed wiring according
to local regulations.
WARNING
Electric shock
Death or serious
personal injury
- Comply with local
regulations as to cable
cross-sections.
DANGER
Electric shock
- Death or serious
personal injury
- Make sure that the
wires to be connected
to the connection
groups below are
separated from each
other by reinforced
insulation in their entire
lengths.
WARNING
Hot liquid
Death or serious
personal injury
- If the temperature is
higher than 50 °C, use
gloves and glasses
when venting the pump
to avoid risk of
scalding.
WARNING
Biological hazard
Death or serious
personal injury
- Flush the pump with
clean water before first
use.