DUAL DISPLAY DIGITAL MULTIMETER 
USER MANUAL 
    
 
 
16
5. MEASUREMENT TUTORIAL 
 
5-1. Voltage measurements (DCV, ACV, DCmV, ACmV) 
1).Press the button to select desired function. 
2).Press [▲]or[▼]to the desired range (if you have no idea about the 
value of input, we suggest you always start at the highest range). 
Press [AUTO/MAN] button for manual or auto-ranging selection. 
3).Connect the test lead to the V and COM input terminals of the 
instrument. 
4).Connect the test lead to the measuring points and read the displayed 
value. 
NOTE: After measuring high voltage to 1000V dc, errors may 
occur when the 100μV is measured.  Allow up to one 
minute prior to making low-level measurements 
 
5-2. Current measurements (DCA, DC 20A, ACA, AC 20A) 
1).Press the button to select function. 
2).Press [▲]or[▼]to the desired range (if you have no idea about the 
value of input. we suggest you always start at the highest 
range).Press [AUTO/MAN] button to change manual or 
auto-ranging. 
3).Connect the test lead to the 2A or 20A and COM input terminals of 
the instrument. 
4).Connect the test lead to the measuring points and read the displayed 
value. 
 
 
DUAL DISPLAY DIGITAL MULTIMETER 
USER MANUAL 
    
 
 
17
5-3. Resistance, capacitance, continuity beeper measurements   
1). Press the button to select function. 
2). Press [▲]or[▼]to the desired range. Press [AUTO/MAN] button 
to change manual or auto-ranging. 
3). Connect the test lead to the  Ω,   and COM input terminals of 
the instrument. 
4). Connect the test lead to the measuring points and read the 
displayed value. 
5-4. Diode test measurements   
1). Press the button to select function. 
2). Connect the test lead to the 
 and COM input terminals of 
the instrument. 
3). Connect the test lead to the semiconductor junction (diode or 
transistor) as shown in Figure 5-1,and read the displayed value. 
z Figure 5-1 
 
Connect to                  Connect to 
“
”           “COM” 
terminal.                      terminal 
 
 
5-5. dBm measurements   
This function converts a voltage measurement into dBm. The function 
can be selected only when a voltage function (volts ac, volts dc, or 
volts ac+dc) is selected. Press [dBm] button, the secondary display 
shows the dBm value that reposed on the voltage value showed in the 
primary display.