DUAL DISPLAY DIGITAL MULTIMETER
USER MANUAL
22
6. MEASUREMENT TECHNIQUES
6-1. dBm measurement technique
dBm is defined as above or below a 1mW reference. A voltage
measurement is converted to dBm using the following formula:
dBm=10*log
10
(1000*voltage value
2
/reference impedance.)
The reference impedance can be set.
For example, the reference impedance of 600Ω, 0.7746V will be
convert to 0 dBm.
6-2. True rms measurement
The true rms (root-mean-square) value of a waveform is equivalent to
dc value that causes the same amount of heat to be dissipated in a
resistor.
Since average-responding meters have been in use for so long, you
may have accumulated test or reference data based on them. Figure
6-1 illustrates the relationship between ac and dc components for
common waveforms, and compares readings for true rms meters and
average-responding meters. Figure 6-1 will help you convert between
the two measurement methods.
DUAL DISPLAY DIGITAL MULTIMETER
USER MANUAL
23
z Figure 6-1: Voltage Conversion