DUAL DISPLAY DIGITAL MULTIMETER 
USER MANUAL 
    
 
 
6 
3. SPECIFICATIONS 
The specifications are operated under the essential conditions as follows: 
z  A 1-year calibration cycle. 
z  An operating temperature of 18 to 28℃ (64.4 to 82.4℉). 
z  Relative humidity not exceeding 90%. 
z  Accuracy is expressed as  ±(percentage of reading + digits). 
z  The AC specification is based on the 50% of duty cycle. 
 
1. DC VOLTAGE OR DCV OF RIPPLE FUNCTION 
RANGE 
RESOLUTION ACCURACY 
500mV 
10μV 
5V 
100μV 
50V  1mV 
500V  10mV 
1200V  100mV 
0.02%+4 
DCV Input Impedance 
10MΩ  in parallel with <100pF, all ranges. 
DCV of Ripple Function  5V~500V 3 ranges. 
DCV of Ripple Input 
Impedance 
10MΩ∕∕    10MΩ  in parallel with <100pF. 
Normal Mode Rejection Ratio >60dB at 60Hz or 50Hz. 
Common Mode Rejection Ratio >90dB at 60Hz or 50Hz. 
Common Mode Voltage 
(Maximum) 
500V dc or peak ac. 
Maximum Input 
450V dc or peak ac continuous on 500mV range an
ripple function. 
1200V dc or peak ac continuous on other range. 
When the input exceeds the full scale of the selected range, the display will appear 
“—OL—“ of over-range indication. 
 
DUAL DISPLAY DIGITAL MULTIMETER 
USER MANUAL 
    
 
 
7
 
2. TRUE RMS AC, AC+DC VOLTAGE   
Accuracy 
between 2% of range and full range. 
RANGE
20Hz-
50Hz 
50Hz-
2kHz 
2kHz-
10kHz
10kHz-
20kHz
20kHz- 
50kHz
50kHz- 
100kHz 
500mV
5V 
50V 
2%+20 5%+50 
500V 
0.5%+50
-------- -------- 
1000V
1%+10
0.3%+30 0.4%+50
-------- -------- -------- 
Input Impedance   
10MΩ  in parallel with < 100pF,all ranges. 
Maximum Input 
450V dc or peak ac continuous on 500mV range. 
1000Vrms on other range. 
Crest Factor Range
3.0 at full scale.
 
AC+DC all ranges 
±(percentage of reading + 50 counts). 
When the input exceeds the full scale of the selected range, the display will appear 
“—OL—“ of over-range indication.
 
3. ACV OF RIPPLE FUNCTION 
RANGE
20Hz-
50Hz 
50Hz-
2kHz 
2kHz-
10kHz
10kHz-
20kHz
20kHz- 
50kHz
50kHz- 
100kHz 
500mV 1%+10 0.3%+30 0.4%+50 0.5%+50 2%+20 5%+50 
Input Impedance   
10MΩ∕∕    10MΩ  in parallel with < 100pF. 
Maximum Input  450V dc or peak ac continuous. 
Crest Factor Range
3.0 at full scale.
 
When the input exceeds the full scale of the selected range, the display will appea