DUAL DISPLAY DIGITAL MULTIMETER
USER MANUAL
6
3. SPECIFICATIONS
The specifications are operated under the essential conditions as follows:
z A 1-year calibration cycle.
z An operating temperature of 18 to 28 (64.4 to 82.4).
z Relative humidity not exceeding 90%.
z Accuracy is expressed as ±(percentage of reading + digits).
z The AC specification is based on the 50% of duty cycle.
1. DC VOLTAGE OR DCV OF RIPPLE FUNCTION
RANGE
RESOLUTION ACCURACY
500mV
10μV
5V
100μV
50V 1mV
500V 10mV
1200V 100mV
0.02%+4
DCV Input Impedance
10MΩ in parallel with <100pF, all ranges.
DCV of Ripple Function 5V~500V 3 ranges.
DCV of Ripple Input
Impedance
10MΩ∕ 10MΩ in parallel with <100pF.
Normal Mode Rejection Ratio >60dB at 60Hz or 50Hz.
Common Mode Rejection Ratio >90dB at 60Hz or 50Hz.
Common Mode Voltage
(Maximum)
500V dc or peak ac.
Maximum Input
450V dc or peak ac continuous on 500mV range an
d
ripple function.
1200V dc or peak ac continuous on other range.
When the input exceeds the full scale of the selected range, the display will appear
“—OL—“ of over-range indication.
DUAL DISPLAY DIGITAL MULTIMETER
USER MANUAL
7
2. TRUE RMS AC, AC+DC VOLTAGE
Accuracy
between 2% of range and full range.
RANGE
20Hz-
50Hz
50Hz-
2kHz
2kHz-
10kHz
10kHz-
20kHz
20kHz-
50kHz
50kHz-
100kHz
500mV
5V
50V
2%+20 5%+50
500V
0.5%+50
-------- --------
1000V
1%+10
0.3%+30 0.4%+50
-------- -------- --------
Input Impedance
10MΩ in parallel with < 100pF,all ranges.
Maximum Input
450V dc or peak ac continuous on 500mV range.
1000Vrms on other range.
Crest Factor Range
3.0 at full scale.
AC+DC all ranges
±(percentage of reading + 50 counts).
When the input exceeds the full scale of the selected range, the display will appear
“—OL—“ of over-range indication.
3. ACV OF RIPPLE FUNCTION
RANGE
20Hz-
50Hz
50Hz-
2kHz
2kHz-
10kHz
10kHz-
20kHz
20kHz-
50kHz
50kHz-
100kHz
500mV 1%+10 0.3%+30 0.4%+50 0.5%+50 2%+20 5%+50
Input Impedance
10MΩ∕ 10MΩ in parallel with < 100pF.
Maximum Input 450V dc or peak ac continuous.
Crest Factor Range
3.0 at full scale.
When the input exceeds the full scale of the selected range, the display will appea
r
“—OL—“ of over-range indication.