Loading...
GDS-806/810/820/840 Programming Manual
24
*WAI (no query form)
WAI prevents the programming instrument from executing further commands or
queries until all pending operations finish.
Syntax
*WAI
:ACQuire:AVERage
Select the average number of waveform acquisition. The range for averaging is
from 2 to 256 in powers of 2.
Note: Before implement this instrument, please apply “:ACQuire:MODe 2” in
advance!
Syntax
:ACQuire:AVERage {1|2|3|4|5|6|7|8}
:ACQuire:AVERage?
Arguments
1Average number is 2 2Average number is 4
3Average number is 8 4Average number is 16
5Average number is 32 6Average number is 64
7Average number is 128 8Average number is 256
Returns
<NR1>