GDS-806/810/820/840 Programming Manual
52
Check sum: The check sum value is coming from the sum of 500 bytes of template
data. This value is 32 bits unsigned integer in hex (4 bytes, high byte first).
Template data: The effective template data which covers 250 points totally(500
bytes). Each point is two bytes signed integer in hex. The valid range is
from –100 to 100(0x0064 ~ 0xFF9C)
Arguments
<X>→Specify the template number (1|100)
Example
The following is a partial of command file including the command. It’s used for
transmitting template data to location 1:
Note: In this case the position is –50 (0xFFCE) dots. The check sum is
0x000016DA
:TEMPlate:MAX (Available for firmware 2.0 and above only)
Sets and query the number of template used for maximum boundary.
Syntax
:TEMPlate:MAX {1|100}
:TEMPlate:MAX?
Returns
<NR1>
Note: A template can be defined as either MAX or MIN template only.