OPERATION
CH1/CH2 Tracking Series Mode
Background
Tracking series operation doubles the Voltage
capacity of the GPD-X303S series by internally
connecting CH1 (Master) and CH2 (Slave) in series
and combining the output to a single channel. CH1
(Master) controls the combined Voltage output
level.
The following describes two types of
configurations depending on the common ground
usage.
Tracking series without common terminal
Connection
Output rating
0 ~ 60V/0 ~ 3A
1. Press the SER/INDEP key
to activate the tracking
series mode. The key LED
turns on.
2. Connect the load to the front panel terminals,
CH1+ & CH2 (Single supply).