GPD-X303S Series User Manual
3. To turn on the output, press
the output key. The key
LED turns on.
4. The CH2 C.V./C.C. PAR.
indicator turns red,
indicating tracking parallel
(PARA) mode.
5. Press the CH1 key (LED
turns on) and then use the
Voltage and Current knob
to set the output voltage
and current. The CH2
output control is disabled.
By default, the Voltage and
Current knob work in the
coarse mode. To activate the
fine mode, press the knob to
turn the FINE LED on.
(Fine control)
6. For the output level and CV/CC status, refer to
the CH1 meter and indicator.
Voltage level
The CH1 meter reading shows
the output voltage. In the above
case, 20.0V.
Current level
Double the amount of CH1
current meter reading. In the
above case, 2.0A x 2 = 4.0A.