MULTI-OUTPUT POWER SUPPLY 
      USER MANUAL 
 
 
2 
2. SPECIFICATIONS 
2-1. General 
Main Supply (switch selectable)  : 100V/120V/220V ± 10%(230V +10%~-6%) 50/60Hz.
Operation Environment 
:Indoor use. 
Altitude up to 2000m. 
Ambient temperature 0℃ to 40℃. 
Relative humidity 80% (maximum). 
Installation category II. 
Pollution degree 2. 
Storage Temperature & Humidity  : -10℃ to 70℃.70% (maximum). 
Accessories  : Operation Manual  ×1. 
Table 2-1:   
  REPLACED FUSE TYPE RATES INPUT TEST LEAD 
MODEL 
Independent Series Para 100V/120V  220V/230V WATTS VA
Current
≦3A 
4A≦
Current<10A
GPS-2303 
0~30V×2 
0~3A×2 
60V 
3A 
30V 
6A 
T6A 
250V 
T3A  250V  350  450 0  2 
GPS-3303 
0~30V×2 
0~3A×2 
60V 
3A 
30V 
6A 
T6.3A  
250V 
T3.15A 250V 420  550 1  2 
GPS-4303 
0~30V×2 
0~3A×2 
60V 
3A 
30V 
6A 
T6.3A  
250V 
T3.15A 250V 420  550 2  2 
GPS-4302 
0~30V×2 
0~2A×2 
60V 
2A 
30V 
4A 
T5A 
250V 
T3A  250V  320  420 2  2 
GPS-4251 
0~25V×2 
0~0.5A×2 
50V 
0.5A
25V 
1A 
T2.5A 
250V 
T1.25A 250V 150  185 4  0 
 
Dimensions : 255(W) ×145(H)  ×265(D) m/m   
Weight (kg)  : 7.0 kg. 
2-2. Operation Mode   
(1) Independent  : Two independent outputs and 
  CH3: 2.2~5.2V output for GPS-4302/4303, 3~6V 
output for GPS-4251, Fixed 5V for GPS-3303. 
CH4: 8~15V output for GPS-4302/4303/4251. 
Output from 0 to rating volts and 0 to rating amperes. 
(2) Series  : Output from 0 to rating volts at rating amperes each. 
: Output from 0 to double rating volts at rating amperes.
(3) Parallel  : Output from 0 to double rating amperes at rating 
volts. 
2-3.  Constant Voltage Operation 
(1) Output voltage range  : 0 to rating voltage continuously adjustable. 
(2) Regulation  : Line regulation≦0.01% + 3mV. 
: Load regulation≦0.01%+3mV(rating current≦3A). 
: Load regulation≦0.02%+5mV ( rating current>3A). 
MULTI-OUTPUT POWER SUPPLY 
      USER MANUAL 
 
 
3
(3) Recovery time  :≦100μs (50% load change, minimum load 0.5A ). 
(4) Ripple & Noise  :≦1mVrms (5Hz-1MHz ). 
(5) Temperature coefficient  :≦300ppm/℃. 
2-4.  Constant Current Operation 
(1) Output current range  : 0 to rating current continuously adjustable. 
(2) Regulation  : Line regulation≦0.2% + 3mA. 
: Load regulation≦0.2% +3mA. 
(3) Ripple current  :≦3mArms. 
2-5. Tracing Operation 
(1) Parallel Operation   
 Regulation  : Line regulation≦0.01% + 3mV. 
: Load regulation≦0.01%+3mV (rating current≦3A). 
≦0.02%+5mV (rating current >3A). 
(2) Series Operation 
 Regulation  : Line regulation≦0.01%+5mV. 
: Load regulation≦300mV. 
 A. Positive and Negative supply (Fig 5-4) CH2 tracking error ≦0.5%+10mV of the CH1 
(No load, with load add load regulation≦300mV) 
 B. Single supply (Fig. 5-3) 
2-6. Meter 
 A. Display A  : 3 digits panel meter×2 (0.5” Red LED display). 
     Display V  : 3 digits panel meter × 2 (0.5” Green LED display). 
 B. Accuracy  :OUT ON±(0.5% of rdg + 2 digits)   
:OUT OFF±(0.5% of rdg+8 digits)(When the output of 
the GPS-2303 is set to OFF, its voltage meter and 
current meter can not preview) 
 C. Voltmeter  : 99.9V of full scale. 
 D. Ammeter  : 9.99A of full scale. 
2-7.  CH3 Output Specifications 
(1) Regulation  : Line regulation≦5mV, load regulation≦15mV. 
(2) Ripple & Noise  :  ≦2mVrms. 
(3) Outout Voltage range  : GPS-4302/4303: 2.2~5.2V±8% continuous adjustment, 
GPS-3303 fixed 5V±8% continuous adjustment, 
GPS-4251 3~6V±8% continuous adjustment 
(4) Output current  : 3A for GPS-3303&4302, 1A for GPS-4303, 2.5A for 
GPS-4251. 
2-8.  CH4 Output Specifications 
(1) Regulation  : Line regulation≦5mV, load regulation≦10mV. 
(2) Ripple & Noise  :≦2mVrms. 
(3) Outout Voltage range  : GPS-4302/4303/4251: 8~15V±8% continuous 
adjustment.