PSW Series User Manual
64
F-03
1 = CC High Speed Priority
3 = CC Slew Rate Priority
5. Press the Voltage knob to save the
configuration setting. ConF will be
displayed when successful.
Voltage
6. If CC Slew Rate Priority was chosen as the
operating mode, set F-06 (Rising Current Slew
Rate) and F-07 (Falling Current Slew Rate) and
save.
F-06 / F-07
0.01A/s~72.00A/s (PSW 30-36)
0.1A/s~144.0A/s (PSW 30-72)
0.1A/s~216.0A/s (PSW 30-108)
0.01A/s~27.00A/s (PSW 80-13.5)
0.01A/s~54.00A/s (PSW 80-27)
0.01A/s~81.00A/s (PSW 80-40.5)
0.01A/s~14.40A/s (PSW 160-7.2)
0.01A/s~28.80A/s (PSW 160-14.4)
0.01A/s~43.20A/s (PSW 160-21.6)
0.001A/s~9.000A/s (PSW 250-4.5)
0.01A/s~18.00A/s (PSW 250-9)
0.01A/s~27.00A/s (PSW 250-13.5)
0.001A/s~2.880A/s (PSW 800-1.44)
0.001A/s~5.760A/s (PSW 800-2.88)
0.001A/s~8.640A/s (PSW 800-4.32)
7. Press the Function key again to exit
the configuration settings. The
function key light will turn off.