ANALOG CONTROL 
125 
External Voltage Control of Current Output 
 
External voltage control of the current output is 
accomplished using the MIL-26 connector on 
the rear panel. A voltage of 0~10V is used to 
control the full scale current of the instrument, 
where:  
Output current = full scale current × (external 
voltage/10) 
When connecting the external voltage source to 
the MIL connectors, use shielded or twisted 
paired wiring. 
PSWEXT-V
Analog 
connector
+
-
16
5
Output 
Terminal
2 core shielded 
wire or twisted 
pair
 
  Pin16 → EXT-V (-) 
  Pin5 → EXT-V (+) 
  Wire shield → negative (-) output terminal 
www.GlobalTestSupply.com