PERFORMANCE ADJUSTMENT
27
PERFORMANCE
ADJUSTMENT
Overview
Background
Performance adjustment checks that the SPS power
supply is performing at the correct specification
level.
Verification item
x Rating Voltage
x Voltage coarse/fine level
x Rating Current
x Current coarse/fine level
Equipment
Digital Multimeter
x DCV Accuracy < 0.1%
x DCA Accuracy < 0.1%
x DCA range: 32A
x Recommended model: GDM-8145
Philips screw
driver
x < 3mm for adjustment points