Troubleshooting
7.3 Event Codes
Event codes are not displayed on controller and must be downloaded from the event log to
be viewed. Troubleshooting actions are provided in Table 5 on page 28.
High Readings Deionized water turbidity is greater than
0.5 NTU
Clean the instrument.
Access Calibration History for turbidity value of
ultra-filtered water.
Verify the flow is between 200–750 mL/min.
Recalibrate the instrument.
Table 6 Additional Malfunctions Not Recorded in the Event Log (continued)
Symptom Possible Cause Corrective Action
Table 7 Event Log List
Event Event # Data1 Data2 Data3
Bubble Reject Change 0
0 = OFF
1= ON
——
Filter Size Change 1
0 = No avg, 1 = 6 sec
2 = 30 sec, 3 = 60 sec
4 = 90 sec
——
Data Log Interval Change 2
0 = 30 sec, 1 = 1min
2 = 5 min, 3 = 10 min
4 = 15 min
——
Power On 3 1
Calibration 4 Cal Gain Initials
Verification 5 Expected Value Meas Value Initials
Dark Reading Warning 6 Measured Value
Temperature Warning 7 Temperature Value
Voltage Warning 8
16 = +5V high
32 = +5V low
64 = V in high
128 = V in low
——
Lamp Warning 9
1 = lamp V high
2 = lamp V low
4 = lamp I high
8 = lamp I low
——
A/D Fail 10 1
Lamp Fail 11 2
Flash Fail 12 4
Output Mode Change 13
0 = Active, 1 = Hold
2 = Transfer, 4 = Normal
——