EasyManua.ls Logo

Hach 1720E - Table 8 Sensor Modbus Registers

Hach 1720E
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
Appendix A Modbus Register Information
Table 8 Sensor Modbus Registers
Group Name Tag Name
Register
#
Data
Type
Length R/W Units (U) Range
Measurements Cal Gain 40013 float 2 R none 0.5 to 2.0
Verification PF Criteria 40062 Integer 1 R/W % 5 to 10
Measurements Turbidity 40001 Float 2 R NTU 0/100
Diagnostics Temperature 40005 Float 2 R Deg C
Diagnostics Dark Reading 40009 Float 2 R NTU 0/100
Diagnostics Raw Turbidity 40011 Float 2 R NTU 0/100
Diagnostics Lamp Voltage 40018 Float 2 R Volts 0/5
Diagnostics Lamp Current 40020 Float 2 R Amps 0/2.75
Diagnostics Plus 5V 40022 Float 2 R Volts 0/5.5
Diagnostics Voltage In 40024 Float 2 R Volts 0/1443
Setup Software Version 40015 Float 2 R
Setup Bubble Rej 40017 Integer 1 R/W On/Off
Setup DataLog Interval 40026 Integer 1 R/W Sec or Min
30 sec, 1 min,
5min, 10 min,
15 min
Setup Sensor Name 40027 String 6 R/W
Setup Filter Size 40033 Integer 1 R/W sec
no averaging,
6, 30, 60, 90
Setup Sensor Ser Num 40036 String 6 R 12 digits
Setup Output Mode 40042 Integer 1 R 0, 1, 2, 4
Setup Set Resolution 40061 Integer 1 R/W decimal places 4, 3, or 2

Table of Contents