EasyManuals Logo

Hach AS950 Basic Operations

Hach AS950
472 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
10. Push Grab. The pump operates in reverse to purge the intake tube. The pump then operates in
the forward direction and collects the sample volume. The pump operates in reverse to purge the
intake tube.
11. After the sample is collected, compare the volume of the sample in the graduated cylinder with
the sample volume entered.
12. If the collected volume is different from the entered sample volume, enter the volume that was
actually collected. Push Next. The controller is adjusted to accurately measure all programmed
volumes.
13. To verify the calibration, push Verify.
14. To save the calibration, push Finish.
Adjust the sample volume calibration (optional)
For the best sample volume accuracy, adjust the sample volume calibration. When the sample
volume is less than 200 mL, the manufacturer recommends that the sample volume calibration is
adjusted.
1. Select Hardware Setup>Liquid Detect>Status. Make sure that the liquid detector is enabled.
2. For the refrigerated sampler or AWRS sampler, collect three grab samples of the volume that is
specified in the sampling program. Refer to the expanded user manual on the manufacturer's
website.
Note: As an alternative, collect three samples with the sampling program. Refer to step 3.
3. For the portable sampler, collect three samples in one bottle (or three bottles) with the sampling
program. For the best accuracy, collect three samples in one bottle. If necessary, change the
sampling program to immediately collect the samples.
Note: Do not use grab samples to adjust the sample volume calibration of a portable sampler.
4. Pour the three samples into a graduated cylinder to measure the total sample volume.
5. Divide the total sample volume by 3 to get the average sample volume.
6. Calculate the adjustment value:
(target sample volume – average sample volume) ÷ target sample volume × 100 = adjustment
value
Where:
Target sample volume = the volume that is specified in the sampling program.
For example, if 200 mL is the target sample volume and the average sample volume is 202 mL,
the adjustment value is –1%.
7. Push MENU.
8. Push CALIBRATION, then select Volume>User Adjust.
9. Enter the adjustment value (e.g., –1%), then click OK.
10. For the refrigerated sampler or AWRS sampler, collect another grab sample as follows:
a. Push Verify.
b. Enter the sample volume that is specified in the sampling program.
c. Push Grab.
Note: As an alternative, do steps 35 again.
11. For the portable sampler, do steps 35 again.
12. If the sample volume collected is not satisfactory when compared to the sample volume that is
specified in the sampling program, do steps 211 again with five samples.
Divide the total volume by 5 to get the average sample volume.
Calibrate the sample volume—liquid detector disabled
When the liquid detector is disabled, calibrate the sample volume by time. The sample volume is
calibrated for the volume specified in the current sampling program. If the sample volume is changed
in the sampling program, manually calibrate the sample volume again for the new volume.
English
17

Table of Contents

Other manuals for Hach AS950

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach AS950 and is the answer not in the manual?

Hach AS950 Specifications

General IconGeneral
Sample FrequencyProgrammable
TypePortable
Sampling MethodVacuum Pump
Temperature ControlRefrigerated
Bottle Configuration24 x 1L bottles
Power Supply12V DC
CommunicationRS-232
Sampler TypePortable or Stationary
Vertical Lift8.2 m
Enclosure RatingIP67
Number of SamplesUp to 999
Storage Capacity24 samples
Bottle Options1L

Related product manuals