DR 5000 Pour-Thru Cell Kit Installation
Assembly
The 1-inch Pour-Thru cell requires the use of a Pour-Thru cell adapter (Figure 1, item 3).
The 1-cm and 5-cm cells use the multi-cell holder supplied with the DR 5000.
Prepare the tubing and Pour-Thru housing
1. Cut the 182 cm (6 ft) length of clear tubing into two equal lengths of tubing,
approximately 91 cm (3 ft) in length.
2. Feed both ends of the clear tubing through the two holes on the side of the housing,
leaving approximately 20 cm (8 inch) inside the housing (Figure 3).
3. Resting the housing on its left side, remove the rubber fitting if installed. Route the
external tubing through the channels molded into the exterior right side of the housing.
Replace the rubber fitting (Figure 3).
Important Note: Do not kink or bend the tubing sharply.
Figure 2 Pour-Thru cell kit components
1 Stand assembly 3 Glass funnel 5 Pour-Thru cell—1-inch or 1-cm
(5 cm sold separately). The 1-inch
cell is shown.
2 Overflow tube assembly 4 Tubing