4
DR 5000 Pour-Thru Cell Kit Installation
1-inch cell assembly
1. Remove the multi-cell holder from the DR 5000 and replace it with the Pour-Thru cell
adapter (Figure 1, Item 3). Do not screw down the Pour-Thru cell adapter.
2. Place the Pour-Thru module on top of the Pour-Thru cell adapter in the DR 5000.
Make sure the locking screw holes are aligned. Secure the assembly with the two
locking screws.
3. Clean the Pour-Thru cell with a lint-free cloth. Insert the Pour-thru cell with the barb
fittings pointing left.
4. Route the tubing as shown in Figure 4 on page 5, trimming tube length as needed to
achieve smooth curves with no kinks or sharp bends.
5. Assemble the stand and connect the tubing, noting flow direction arrows molded into
the housing where the tubes emerge. The inflow tube is connected to the glass funnel;
the outflow tube is connected to the overflow tube assembly (Figure 5 on page 5).
6. Connect the drain tube (black rubber tube) to the bottom fitting of the overflow tube
assembly. Connect the other end of the drain tube to an appropriate waste vessel.
The tubes can be shortened if necessary.
Figure 3 Tubing installed in housing and channels with rubber fitting
1 Tubing (8” inside housing) 2 Rubber fitting 3 Tubing channels